Please use this identifier to cite or link to this item:
https://hdl.handle.net/2440/38931
Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Halim, Dunant | en |
dc.contributor.author | Petersen, Dick | en |
dc.contributor.author | Cazzolato, Benjamin Seth | en |
dc.date.issued | 2004 | en |
dc.identifier.uri | http://hdl.handle.net/2440/38931 | - |
dc.description.abstract | This report discusses the results from the second calibration test performed on the Phone-Or 3D optical Energy Density probe. The first calibration test found several performance issues associated with high level of noise in the Phone-Or probe. Since then, the Phone-Or manufacturer has addressed those issues by modifying and re-calibrating the probe. To investigate the current performance of modified/re-calibrated probe, a second calibration test was performed in the Anechoic chamber at the University of Adelaide. The test results show that the Phone-Or probe's performance has improved considerably with its self-noise level, particularly due to 50 Hz harmonics, been reduced significantly. | en |
dc.format.extent | 340640 bytes | en |
dc.format.mimetype | application/pdf | en |
dc.language.iso | en | en |
dc.title | The phone-or 3D optical energy density probe: calibration test report of the modified/re-calibrated probe | en |
dc.type | Report | en |
dc.contributor.school | School of Mechanical Engineering | en |
Appears in Collections: | Acoustic, Vibration and Control Group reports Environment Institute publications |
Files in This Item:
File | Size | Format | |
---|---|---|---|
hdl_38931.pdf | 332.66 kB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.